9322 adalah probe diferensial yang digunakan untuk mengukur tegangan tinggi dengan Hioki Memory HiCorders.
Measurement functions | DC mode: Waveform monitor output, DC to 10 MHz ±3 dB AC mode: Detection of power line surge noise, 1 kHz to 10 MHz ±3 dB (Low frequency cut-off frequency 1 kHz ± 300 Hz) RMS mode: Rectified RMS output of DC and AC voltages, DC, 40 Hz to 100 kHz, Response speed: 200 ms or less (400 V AC) |
---|---|
Max. allowable input | 2000 V DC, 1000 V AC |
Max. rated voltage to earth | When using grabber clip 9243: 1500 V AC/DC (CAT II), 600 V AC/DC (CAT III) When using alligator clip: 1000 V AC/DC (CAT II), 600 V AC/DC (CAT III) |
Output | Voltage division ratio: 1/1000, BNC terminal (DC/AC/RMS 3-mode selectable output) |
DC amplitude accuracy | ±1 % f.s. (1000 V DC or less), ±3 % f.s. (2000 V DC or less) (f.s.=2000 V DC) |
RMS amplitude accuracy | ±1 % f.s. (DC, 40 Hz to 1 kHz), ±4 % f.s. (1 kHz to 100 kHz) (f.s.=1000 VAC) |
Input resistance, capacity | H-L: 9 MΩ, approx 10 pF (C at 100 kHz) H-case, L-case: 4.5 MΩ, approx 20 pF (C at 100 kHz) |
Power supply | +5 to +12 V, less than 300 mA. (DC jack OD 5.5 mm (0.22 in), ID 2.1 mm (0.08 in)), (1) Via AC adapter 9418-15, (2) Via Logic terminal on Memory HiCorder through Power cord 9324, (3) Via F/V Unit 8940’s sensor terminal through Power cord 9325, (4) Via DC power output terminal attached to the input unit for the 8855 through Power cord 9328, (5) Via the 8860 series dedicated Probe Power Unit 9687 through Power cord 9248, (6) Via MR6000 dedicated Probe Power Unit Z5021 through Power cord 9248 |
Dimensions and mass | 70 mm (2.76 in)W × 150 mm (5.91 in)H × 25 mm (0.98 in)D, 350 g (12.3 oz), Cord length: Input 46 cm (1.51 ft), Output 1.3 m (4.27 ft) |
Accessories | Alligator clips ×1 (red/black set), Grabber clip 9243 ×1 (red/black set), Carrying case C0203 ×1, Instruction manual ×1 |
■ How to power the 9322 with a Hioki Memory HiCorder (Via MR6000 dedicated Probe Power Unit Z5021)
*3: When a current sensor is connected to the Probe Power Unit Z5021, or the Current Unit 8971, U8977, the number of the 9322 connection including the current sensor is limited (up to a total of 9 including the current sensor).
Main unit | Number of the 9322 connections |
---|---|
MR6000 MR6000-01 |
8 (Combined with 9248 cable) *3 |
■ How to power the 9322 with a Hioki Memory HiCorder (Via MR8875 DC power output terminal)
(Caution) To avoid electric shock, connect the power cord supplied with the main unit (Memory HiCorder or AC adapter for the 9322) to a grounded 2-pole outlet.
Main unit | Number of the 9322 connections |
---|---|
MR8875 | Unavailable |
■ How to power the 9322 with a Hioki Memory HiCorder (Via F/V Unit 8940’s sensor terminal)
(Caution) To avoid electric shock, connect the power cord supplied with the main unit (Memory HiCorder or AC adapter for the 9322) to a grounded 2-pole outlet.
*1: Discontinued model
Main unit | Required cord | Number of the 9322 connections | Number of current sensors that can be used simultaneously with the 9322 |
---|---|---|---|
8860 *1 8860-50 *1 8861 *1 8861-50 *1 |
9325 | 6 | 8 |
8855 *1 | N/A | N/A | N/A |
8841 *1 8842 *1 |
9325 | 6 | 4 |
8835-01 *1 | 9325 | 4 | 4 |
8835 *1 | Unavailable | Unavailable | Unavailable |
8826 *1 | 93256 | 6 | 6 |
■ How to power the 9322 with a Hioki Memory HiCorder (Via Logic terminal on Memory HiCorder)
(Caution) To avoid electric shock, connect the power cord supplied with the main unit (Memory HiCorder or AC adapter for the 9322) to a grounded 2-pole outlet.
*1: Discontinued model
Main unit | Required cable | Number of the 9322 connections | Number of logic probes that can be used simultaneously with the 9322 |
---|---|---|---|
MR6000 MR6000-01 |
N/A | N/A | N/A |
MR8740T | N/A | N/A | N/A |
MR8740 MR8741 |
9323 + 9324 *1 | 4 | 9322 ×4: Unavailable 9322 ×3: Unavailable 9322 ×2: Unavailable 9322 ×1: 2 |
MR8827 | 9323 + 9324 *1 | 4 | 9322 ×4: Unavailable 9322 ×3: Unavailable 9322 ×2: Unavailable 9322 ×1: 2 |
8847 *1 MR8847-01 *1 MR8847-02 *1 MR8847-03 *1 MR8847-51 MR8847-52 MR8847-53 |
9323 + 9324 *1 | 4 | 9322 ×4: Unavailable 9322 ×3: Unavailable 9322 ×2: Unavailable 9322 ×1: 2 |
8870-20 *1 MR8870-20 |
Unavailable | ||
MR8875 | Unavailable | ||
MR8880-20 | Unavailable | ||
8860 *1 8860-50 *1 8861 *1 8861-50 *1 |
9323 + 9324 *1 | 2 | 9322 ×2: Unavailable 9322 ×1: 3 |
8855 *1 | 9323 + 9324 *1 | 2 | 9322 ×2: Unavailable 9322 ×1: 3 |
8841 *1 8842 *1 |
9324 *1 | 1 | Unavailable |
8835-01 *1 | 9324 *1 | 1 | Unavailable |
8835 *1 | 9324 *1 | 1 | Unavailable |
8826 *1 | 9324 *1 | 4 | 9322 ×4: Unavailable 9322 ×3: 4 9322 ×2: 6 9322 ×1: 7 |
■ How to power the 9322 with a Hioki Memory HiCorder (Via the 8860 series dedicated Probe Power Unit 9687)
(Caution) To avoid electric shock, connect the power cord supplied with the main unit (Memory HiCorder or AC adapter for the 9322) to a grounded 2-pole outlet.
*1: Discontinued model
*2: The number of 9322 that can be connected varies depending on the combination with the clamp probe connected to the probe power supply unit 9687.
Main unit | Number of the 9322 connections |
---|---|
8860 *1 8860-50 *1 8861 *1 8861-50 *1 |
8 (Combined with 9248 cable) *2 |
■ How to power the 9322 with a Hioki Memory HiCorder (Via DC power output terminal attached to the input unit 8950, 8952, 8953-10, 8955 for the 8855)
(Caution) To avoid electric shock, connect the power cord supplied with the main unit (Memory HiCorder or AC adapter for the 9322) to a grounded 2-pole outlet.
*1: Discontinued model
Main unit | Required cable |
---|---|
8855 *1 | 9328 |
Reviews
There are no reviews yet.